PXD222-lecroy-replacement-zt44211 The ZTEC ZT4211 PXI is a modern oscilloscope that provides extensive functionality, in most cases beyond that of the PXD222. See below for a list of the differences between the ZT4211 and the PXD222. Because the instrument functionality, command set, and I/O ranges are different, there is some effort involved with the integration of the ZT4211 in place of the PXD222. ZTEC's application engineering team has significant experience with the functional differences between the PXD222 and the ZT4211, as well as the application details of using oscilloscopes in ATE test programs. This experience is invaluable when assisting a test system integrator in successfully modifying an existing ATE system or TPS.

Please, visit the ZTEC ZT4210 Series page for more information on the ZT4211.



Specification PXD222 ZT4211PXI
Input Channels 2 2 (ZT4211)
or 4 (ZT4212: VXI/LXI only)
Platform & Size 2 slot 3U PXI Single-Wide 3U PXI
(PCI, VXI, & LXI available)
Digitizer Resolution 8 bits 8 bits
Input Bandwidth 200 MHz (-3 dB) 300 MHz
Input Coupling DC, AC DC, AC
Input Filter Not supported 20 MHz or Bypass
Filter Stopband Rejection: approximately 3dB @ 20 MHz
Input Impedance 1MΩ(±1%)//15 pF (±2 pF) 1MΩ(±1%) || 12 pF or 50Ω
Maximum Input direct(1:1) 300 V CAT II
with 10:1 probe 1000V
50Ω: ±5 V (DC + peak AC), CAT II
  Input load protection @ ±6 VDC
1MΩ: ±300 V [DC + peak AC (<100 kHz)], CAT II
  Peak AC, de-rated 20 dB/decade above 100 kHz
Sample Rate Glitch Capture Mode: up to 2.5 GS/s
Scope Record Mode: up to 20 MS/s
10 kS/s to 500 MS/s real-time,
1 GS/s Interleaved to 100 GS/s Interleaved Equivalent Time
Maximum Record Length Glitch Capture Mode: 250 points
Scope Record Mode: 27K points (up to 20 MS/s)
128 MiSamples or 256 MiSamples Interleaved
Trigger Modes Edge, Video & Pulse Width
Video: PAL, PAL+, NTSC, SECAM, Filed 1 or Field 2, Line Select
Edge, Pattern, State, Pulse Width, Video
Video: PAL (50 Hz), NTSC (60 Hz), SECAM (50 Hz)
Standard, Field, Line selectable
Pre-Trigger Capture 90% of the recorded time 0 to 100% of acquisition window
Trigger Sources Ch1,Ch2,PXI_TRIG[n],PXI_STAR[n] Channels 1 to 2, Channels 3 to 4 (ZT4212), External Input, Pattern, Software, TTL Trigger 0-7, Star Trigger (PXI/PCI), ECL Trigger 0-1 (VXI)
Trigger Holdoff Range Up to 100 times the recorded span time
Maximum Delay: 10 seconds
Programmable Delay: 0 to 100 seconds
Arm Not supported Trigger qualifier
Clock Accuracy ± 100 ppm < 2.5 ppm
External Out Not supported Trigger Out, 10 MHz, Event Out,
Probe Compensation, Programmable Clock/Pulse
Power +5 VDC   1.6 A (8 W) 2 Channels PXI:
Typical 17.0 W
Max. 24.0 W

support_sm_whiteGet your questions answered thoroughly and quickly by submitting a ticket to the ZTEC Customer Support Center.