ZT4628 VXI
Analogic DBS907AB + DBS9905 Replacement Solution
The ZTEC ZT4628 is a modern
oscilloscope that provides extensive functionality, in most cases beyond that of the DBS907AB.
See below for a list of the differences between the ZT4628 and the DBS907AB. Because the instrument functionality, command set, and I/O
ranges are different, there is some effort involved with the integration of the ZT4628 in place of the DBS907AB.
ZTEC's application engineering team has significant experience with the functional differences between the DBS907ABA
and the ZT4628, as well as the application details of using oscilloscopes in ATE test programs. This experience
is invaluable when assisting a test system integrator in successfully modifying an existing ATE system or TPS.
Please, visit the ZTEC ZT4620 Series
page for more information on the ZT4628.
| Specification | DBS907AB + DBS9905 | ZT4628VXI |
|---|---|---|
| Input Channels | 1 or 2 | 2 (ZT4628) or 4 (ZT4629) |
| Input Bandwidth
(-3dB, 50Ω) |
250 MHz on 50 mV range,
500 MHz on other ranges |
500 MHz all ranges |
| Input Impedance | 50Ω or 1MΩ | 50Ω or 1MΩ |
| Input Coupling | DC or AC | DC, AC or LF Reject |
| Input Filter | Not supported | 30 MHz HF Reject |
| Maximum Voltage | 1MΩ: ±250 V (DC + peak AC)
50Ω: ±5 VDC or 5 Vrms |
1MΩ: ±250 V (DC + peak AC)
50Ω: ±5 VDC or 5 Vrms Overload protection switch |
| Voltage Ranges (p-p) | 10 ranges:
50 mV, 100 mV, 200 mV, 500 mV, 1V, 2V, 5V, 10V, 20V, 50V |
6250 ranges:
8 mV to 50 V in 8 mV steps |
| Offset Adjustment | ±2V for ≤ 500 mV FSR,
±20V for ≤ 5 V FSR, ±200V for ≤ 50 V FSR |
±2V for ≤ 400 mV FSR,
±10V for ≤ 2 V FSR, ±50V for ≤ 10 V FSR, ±250V for ≤ 50 V FSR |
| DC Accuracy | < ±(2% FSR + 0.4% offset) | < ±(1% FSR + 0.5% offset +
1 mV) |
| Offset Drift | < ± 0.1% FSR / °C | < ± 0.1% FSR / °C |
| Gain Drift | < ± 0.02% FSR / °C | Not specified |
| ADC Resolution | 8 bit | 8-bit,
Up to 32-bit with averaging |
| Sample Rate | 100 S/s to 1 GS/s | 10 S/s to 1 GS/s real-time,
2.5 GS/s to 100 GS/s ETS |
| Maximum Memory | 2M or 8M samples | 256M samples |
| Segmented Memory | Up to 8k segments | Up to 32k segments |
| Pre-Trigger Delay | Up to 8M samples | Up to 256M samples |
| Post-Trigger Delay | Up to 200M samples | Up to 256M samples + 200 s |
| Trigger Modes | Edge, Video (TV) | Edge, Video (TV), Pulse Width,
Pattern, Glitch, A-B |
| Trigger Holdoff | 100 ns to 6.5 ms | 0 ns to 100 s |
| Trigger Event
Counter |
Not Supported | Up to 65,536 events |
| Trigger Sources | Inputs, External | Inputs, External, TTLTRGn*,
ECLTRGn |
| Trigger Coupling | DC or AC | DC or Noise Reject |
| External Trigger In | ±3V threshold,
50 Ω or 1 MΩ, 250 MHz bandwidth |
±2V threshold,
50 Ω or 1 MΩ, 250 MHz bandwidth |
| External Clock In | 10 MHz or 500 MHz,
±3V threshold |
10 MHz,
±2V threshold |
| Arm | Not supported | Trigger qualifier |
| Clock Accuracy | < 25 ppm | < 2.5 ppm |
| External Out | Not supported | Trigger Out, 10 MHz, Event
Out, Probe Compensation, Programmable Clock/Pulse |
| Power | 2 Channels: 33W | 2 Channels: 22.5W |
| Platform | VXI | VXI, PXI, PCI, LXI |
Get your questions answered thoroughly and quickly by submitting a ticket to the ZTEC Customer Support Center.