ZT452-21 VXI
Agilent (HP) E1426A Replacement Solution
The ZTEC ZT452VXI-21 is a modern VXI oscilloscope that provides extensive functionality, in most cases beyond that of the E1426A.
The main difference between the two instruments is that the Agilent E1426A is a low-sample rate oscilloscope that uses repetitive sampling to
reconstruct high-bandwidth waveforms. Whereas the ZTEC ZT452-21 provides the same reconstructive repetitive sampling, it also provides single-shot
acquisition up 1GS/s. See below for a list of the differences between the ZT452-21 and the E1426A. Note that there are no differences in other
specifications not listed in the table. Because the instrument functionality, command set, and I/O ranges are different, there is more effort
involved with the integration of the ZT452-21 in place of the E1426A. ZTEC's application engineering team has significant experience with the
functional differences between the E1426A and the ZT452-21, as well as the application details of using VXI oscilloscopes in ATE test programs.
This experience is invaluable when assisting a test system integrator in successfully modifying an existing ATE system or TPS.
| Specification | E1426A | ZT452-21 VXI |
|---|---|---|
| Physical Width | 2 slots | 1 slot |
| Maximum Sample Rate | 20 MS/s | 1 GS/s |
| Maximum Memory | 1,024 points | 16,777,216 points |
| Maximum Sweep Time | 50 s | 6,710 s |
| Input Coupling | DC, AC, AC-LFR, LPF 30MHz | DC, AC, LPF 20MHz |
| Maximum Input | 250 VDC (1 MΩ)
5 Vrms (50 Ω) |
210 VDC (1 MΩ)
5 VDC (50 Ω) |
| Voltage Ranges | 0.008 to 40 Vpp
5000 steps |
0.5 to 100 Vpp (1 MΩ)
0.05 to 10 Vpp (50 Ω) 8 steps |
| DC Offset Adjust | ±2V (0.008 to 0.4Vpp)
±10V (0.4 to 2Vpp) ±50V (2 to 10Vpp) ±250V (10 to 40Vpp) |
±5V (0.5 to 5Vpp, 1 MΩ)
±100V (10 to 100Vpp, 1 MΩ) ±0.5V (0.05 to 0.5Vpp, 50 Ω) ±10V (1 to 10Vpp, 50 Ω) |
| DC Gain Accuracy | 1.25% | 1.5% |
| DC Offset Accuracy | 0.5% + 2% of range | 1% + 2% of range |
| External In/Out | AC Cal Out, DC Cal Out,
Trigger Out |
Arm In, Clock In, Trigger In,
Ref Out (DC Cal, AC Cal, Trigger, Arm, 10 MHz, Pulse) |
Get your questions answered thoroughly and quickly by submitting a ticket to the ZTEC Customer Support Center.